AUTO REF/KERAUTOMETER RMK-200                                               CE

 

 

 

 

Specifications

 

 

Refractometry

 

 

Vertex Distance (VD)

 

 

0.0,10.0,12.0,13.5,15.0mm

 

 

Sphere Power (SPH)

 

 

-20.00 ~ +20.00D (at the vertex distance of 12mm)

 

(Increments selectable between 0.12D and 0.25D)

 

 

Cylinder Power (CYL)

 

 

0.00 ~ +10.00 D

 

(Increments selectable between 0.12Dand 0.25D)

 

 

Axis (AX)

 

 

1 ~ 180° (Increments: 1°)

 

 

Cylinder Form

 

 

-, +, MIX

 

 

Pupil Distance (PD)

 

 

10 ~ 85mm

 

 

Minimum Pupil Diameter

 

 

2.0mm

 

 

Keratometry

 

 

Radius of Curvature

 

 

5.0 ~ 10.2mm (Increments: 0.01mm)

 

 

Corneal Power

 

 

33.00 ~ 67.50D

 

(Increments selectable between 0.12D and 0.25D)

 

 

Corneal astigmatism

 

 

0.00~-15.00D

 

(Increments selectable between 0.12D and 0.25D)

 

 

Axis

 

 

1 ~ 180° (Increments: 1°)

 

 

Corneal Diameter

 

 

2.0 ~ 12.0 mm (Increments: 0.1mm)

 

 

Memory of Data

 

 

10 measured value for each right and left eye

 

 

Internal Printer

 

 

Uses thermal line printer

 

 

Monitor

 

 

5.6-inch color TFT LCD monitor

 

 

Environmental requirements:

 

 

Operation

 

 

Temperature: + 10 to + 40°C

 

Humidity: 30 to 85% RH

 

 

Storage and Transportation

 

 

Temperature: - 10 to + 55°C

 

Humidity: 10 to 95% RH

 

 

Power supply

 

 

AC220V, 50Hz

 

 

Size

 

 

280(W) × 475(D) ×475(H)mm

 

 

Weight

 

 

Approximately 26kg

 